Assessing quality of OCT scans impacts clinical-decision process
With imaging artifacts common, clinicians must be focused on key points
Many ocular factors contribute to the incidence of artifacts. Myopic patients are not represented in the normative database even though it is an important cause of errors or misinterpretation.
Reviewed by David S. Greenfield, MD
An assessment of the quality of
A number of ocular factors contribute to the incidence of artifacts, a frequent one being that highly myopic patients are not represented in the normative database despite the fact that high myopia is known to be an important cause of errors or misinterpretation, Dr. Greenfield said. Other factors to watch for include significant ocular surface disease, opacification in the lens or vitreous, and peripapillary atrophy.
Four parts of the scan to study closely are signal strength (refer to the manufacturer’s specifications); the thickness map (look for the presence or absence of shadowing artifacts); the deviation map (has the optic disk been automatically and accurately delineated, and can any subtle features related to motion artifacts be detected); and individual tomograms (search for signs of segmentation failure).
Monitoring the recommended individual signal-to-noise ratios is clinically relevant because a reduction in the signal strength will not only affect the estimation of retinal fiber layer (RNFL) thickness, but is also highly associated with an increase in the measurement’s variability and failure of the measurement algorithm (as demonstrated in the figure).
Details in the imaging
RNFL imaging artifacts can be very subtle, arising from causes such as a horizontal translation of the blood vessels due to eye movement during imaging acquisition. Other common problems are blink artifacts and shadowing artifacts from vitreous opacities.
Eyes with severe pathologic myopia often have what Dr. Greenfield refers to as the “trifecta,” an incorrect axial alignment with shadowing artifacts, poor delineation of the optic disc border, and clear evidence of segmentation failure, all contributing to RNFL measurements that are “off the chart.”
Macular artifacts are even more common than RNFL artifacts and can be found in eyes with pathology on the surface of the macula, intraretinal edema, and even associated with subretinal pathology such as choroidal neovascularization, confounding the GCIPL measurement algorithm. Extremely large regions of macular pathology can also adversely affect the RNFL thickness algorithm and resemble glaucomatous RNFL atrophy.
RELATED:
The GCIPL thickness had been increasing over time due to worsening CME, masking the ability to detect progressive glaucoma. Serial examination of the RNFL images showed progressive atrophy in the average, superior, and inferior RNFL thickness that was not detectable in the macular region because of the presence of CME.
“Poor quality scans with some measures may not preclude high quality scans with other measures in different anatomic regions,” Dr. Greenfield said.
In terms of disease detection, many normal eyes with high axial myopia have anomalous tilted discs making it difficult to clinically establish the presence of glaucomatous damage, he noted. Close examination may show that the neural rim of the optic disc is intact, and the visual field may show a characteristic enlargement of the physiologic blind spot, he added.
The false identification of glaucoma using OCT in such eyes
Dr. Greenfield managed the case of a glaucoma patient with a reproducible paracentral visual field scotoma, normal RNFL thickness, and localized atrophy in GCIPL thickness due to centralized glaucoma damage in the macular region. Such cases remind clinicians of the importance of examining the central 10-2 visual field and GCIPL thickness in addition to performing RNFL imaging in eyes with glaucoma, he said.
A final factor that must be considered when analyzing OCT scans over time is the impact of the aging process, Dr. Greenfield said, adding that serial imaging over an extended period of time may demonstrate slow changes in the RNFL thickness due to normal aging. He urged caution when interpreting eyes with very slowly progressive, isolated changes in RNFL thickness with stable IOP and stable visual function.
RELATED:
Disclosures:
David S. Greenfield, MD
P: 516/515-1500 E: [email protected]
This article was adapted from Dr. Greenfield’s presentation at the 2019 meeting of the American Glaucoma Society. He did not report any relevant financial disclosures.